Show simple item record

contributor authorW. G. Deskins
contributor authorD. C. Winter
contributor authorH.-P. Sheng
contributor authorC. Garza
date accessioned2017-05-08T23:19:37Z
date available2017-05-08T23:19:37Z
date copyrightNovember, 1985
date issued1985
identifier issn0148-0731
identifier otherJBENDY-25807#304_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/99497
description abstractAn acoustic plethysmograph intended to measure the body volume of premature infants has been developed using the principle of the Helmholtz resonator, in which the resonance frequency is dependent on the volume of the resonating cavity. A prototype system was built and used to measure the volume of inanimate objects and newborn miniature pigs. Results for inanimate objects agree within 1 percent with comparable measurements by water displacement. Results of the animal body volume measurements compare favorably (within an average of 1.1 percent) with those obtained using hydrostatic weighing.
publisherThe American Society of Mechanical Engineers (ASME)
titleAn Acoustic Plethysmograph to Measure Total Infant Body Volume
typeJournal Paper
journal volume107
journal issue4
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.3138560
journal fristpage304
journal lastpage308
identifier eissn1528-8951
keywordsAcoustics
keywordsEngineering prototypes
keywordsCavities
keywordsDisplacement
keywordsVolume measurement
keywordsWater
keywordsResonance
keywordsHydrostatics AND Measurement
treeJournal of Biomechanical Engineering:;1985:;volume( 107 ):;issue: 004