Show simple item record

contributor authorHuxman, Connor
contributor authorUpdegrove, Gary
contributor authorArmstrong, April
contributor authorWee, Hwabok
contributor authorFrecker, Mary I.
contributor authorButler, Jared
contributor authorLewis, Gregory S.
date accessioned2026-08-23T08:31:24Z
date available2026-08-23T08:31:24Z
date copyright2026/01/01
date issued2026
identifier issn0148-0731
identifier otherbio-25-1157.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4316673
description abstractAbstract. Axial interfragmentary motion is known to stimulate fracture healing. A mechanically compliant fracture fixation plate incorporating flexures is proposed to provide controlled axial micromotion to long bone fractures. To explore the concept's feasibility, computational modeling of general diaphyseal and distal femur fractures treated with both rigid and compliant plates is conducted. In Part I of this study, a diaphyseal fracture finite element model for novel compliant plates is validated against experimental data with good agreement. In Part II, a parametric analysis is conducted using the validated model to characterize the performance of many compliant plate designs with varying geometry and materials. Under axial loading, all compliant plate configurations provided greater (1.03 mm versus 0.22 mm) and more symmetric (270–390%) axial interfragmentary motion than rigid plates. Steel compliant plates with thicker flexures (0.3–0.6 mm) may provide the best performance given their enhanced motion and comparable bending/torsional rigidity. In Part III, compliant plates are adapted for use in treating distal femur fractures. Results demonstrate that compared to a rigid plate, a compliant distal femur plate with increased thickness can effectively modulate interfragmentary motion—that is, increase the insufficient near cortex motion under low loads (from 0.14 mm to 0.23 mm) and reduce the excessive far cortex motion under large loads (from 7.96 mm to 2.54 mm). Flexure-based locking plates represent a promising new approach to treating diaphyseal and/or distal femur fractures. Additional research is needed to investigate in vivo performance.
publisherThe American Society of Mechanical Engineers (ASME)
titleFlexure-Based Locking Plates Can Modulate Interfragmentary Motion in Distal Femur and Diaphyseal Fractures: A Parametric Finite Element Analysis
typeJournal Paper
journal volume148
journal issue1
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4070406
journal fristpage650
journal lastpage655
page6
treeJournal of Biomechanical Engineering:;2026:;volume( 148 ):;issue:001
contenttypeFulltext


Files in this item