description abstractWith advances in biotechnology, the field of cryopreservation has been continuously developed and improved. Typical cryo-container was designed with minimal flow to avoid possible structural defects in LN2 tank, which has a higher thermal conductivity than vapor nitrogen tank. If cells are placed in typical cryo-container and stored in VN2 tank, cross-contamination can be prevented, but the cell viability after thawing may be reduced. The structure of typical cryo-containers is not optimized for vaporized nitrogen to flow quickly into the container and its circulation well. Therefore, we proposed new cryo-container models that can maintain mechanical strength while optimizing the fluid flow structure, and performed thermal–structural coupled field analysis on cryo-containers. We confirmed the cryo-containers by comparing the equivalent stress distributions formed around through holes and evaluating thermal equilibrium in the cryogenic steady-state through flow analysis. Prototype cryo-containers and typical cryo-containers were placed in VN2 tank for a period of time to observe temperature changes. As a result, the time it takes to reach the temperature equilibrium has been reduced to 55% level compared with the typical cryo-containers. Additionally, C2C12 and hADMSC cells were checked after storage under two temperature conditions (−80 and −196 °C). In both cell, viability, adhesion, and relative cell proliferation were improved by up to 15–20% in new containers compared to typical products. The developed container is expected to maintain stability well by being applied to storage and transportation of advanced medicines that require cryopreservation.
publisherThe American Society of Mechanical Engineers (ASME)
titleComparison of Temperature Equilibrium Rate and Cell Growth/Viability Under Air Circulation in Cryogenic Storage Container
typeJournal Paper
journal volume16
journal issue4
journal titleJournal of Medical Devices
identifier doi10.1115/1.4054833
journal fristpage41003-1
journal lastpage41003-13
page13
treeJournal of Medical Devices:;2022:;volume( 016 ):;issue: 004
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record