Show simple item record

contributor authorScott, Joel C. R.
contributor authorDoman, Darrel A.
contributor authorJohnston, Clifton R.
date accessioned2022-02-04T22:05:25Z
date available2022-02-04T22:05:25Z
date copyright10/1/2020 12:00:00 AM
date issued2020
identifier issn1932-6181
identifier otherbio_142_11_111014.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4274851
description abstractStents were manufactured to investigate the impact of altering stent design characteristics (leg length, bend angle, bend radius, and wire diameter) on radial force generation. Results from this design parameter study showed that leg length, bend angle, and wire diameter have a statistically significant impact on radial force generation, while lesser changes in bend radius did not (1.00 mm versus the original 0.794 mm). However, a larger variation of this parameter (1.588 mm) was statistically significant. Results gathered for all parameters were used in the creation of a prototypal software. Using input values of patient-specific arterial diameter and compliance, as well as stent design characteristic dimensions, this program has been developed to predict stent radial force at varying levels of expansion.
publisherThe American Society of Mechanical Engineers (ASME)
titleA Parametric Analysis of Endovascular Stent Geometry Manipulation on Radial Force Performance
typeJournal Paper
journal volume14
journal issue4
journal titleJournal of Medical Devices
identifier doi10.1115/1.4048233
journal fristpage041004-1
journal lastpage041004-11
page11
treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 004
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record