Show simple item record

contributor authorOosting, R. M.
contributor authorOuweltjes, K.
contributor authorHoeboer, M. D. B.
contributor authorHesselink, L.
contributor authorMadete, J. K.
contributor authorDiehl, J. C.
contributor authorGroen, R. S.
contributor authorWauben, L. S.G. L.
contributor authorDankelman, J.
date accessioned2022-02-04T14:43:46Z
date available2022-02-04T14:43:46Z
date copyright2020/02/05/
date issued2020
identifier issn1932-6181
identifier othermed_014_01_011106.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4274251
description abstractTo comply with the large global need for surgery, surgical equipment that fits the challenging environment in low- and middle-income countries (LMICs) should be designed. The aim of this study is to present a context-specific design of an electrosurgical unit (ESU) and a monopolar handheld to improve global access to surgery. This paper presents both a detailed description of electrosurgery in clinical practice in LMICs and the design of an ESU generator and monopolar handheld for this specific setting. Extensive fieldwork (by means of surveys, interviews, observations, and collection of maintenance records) was done by authors RO, KO, and LH. Feedback from users working in Kenya on the first demonstrator designs was obtained, after which the designs were adapted into conceptual prototypes. These were further evaluated by surveying respondents who attended the annual meeting of the College of Surgeons of East, Central, and Southern Africa (COSECSA) in Kigali, Rwanda in December 2018. Conceptual prototypes were developed for (a) an affordable ESU that is compact and battery powered and (b) a robust reusable monopolar handheld that can be cleaned in the autoclave and by chemicals (e.g., glutaraldehyde solution). The conceptual prototypes were positively received by the 51 respondents of the survey. The findings from the field work and the feedback from users during the design phase have led to a clear understanding of the specific needs and potential solutions. The presented conceptual prototypes need to be further developed into functional prototypes, which could be implemented in Kenya and other settings for further evaluation.
publisherThe American Society of Mechanical Engineers (ASME)
titleA Context-Specific Design of an Electrosurgical Unit and Monopolar Handheld to Enhance Global Access to Surgical Care: A Design Approach Based on Contextual Factors
typeJournal Paper
journal volume14
journal issue1
journal titleJournal of Medical Devices
identifier doi10.1115/1.4045966
page11106
treeJournal of Medical Devices:;2020:;volume( 014 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record