Show simple item record

contributor authorTilton, Maryam
contributor authorArmstrong, April D.
contributor authorWee, Hwabok
contributor authorHast, Michael W.
contributor authorManogharan, Guha
contributor authorLewis, Gregory S.
date accessioned2022-02-04T14:18:42Z
date available2022-02-04T14:18:42Z
date copyright2020/03/27/
date issued2020
identifier issn0148-0731
identifier otherbio_142_08_081005.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4273404
description abstractInternal fixation with the use of locking plates is the standard surgical treatment for proximal humerus fractures, one of the most common fractures in the elderly. Screw cut-out through weak cancellous bone of the humeral head, which ultimately results in collapse of the fixed fracture, is the leading cause of failure and revision surgery. In an attempt to address this problem, surgeons often attach the plate with as many locking screws as possible into the proximal fragment. It is not thoroughly understood which screws and screw combinations play the most critical roles in fixation stability. This study conducted a detailed finite element analysis to evaluate critical parameters associated with screw cut-out failure. Several clinically relevant screw configurations and fracture gap sizes were modeled. Findings demonstrate that in perfectly reduced fracture cases, variation of the screw configurations had minor influence on mechanical stability of the fixation. The effects of screw configurations became substantial with the existence of a fracture gap. Interestingly, the use of a single anterior calcar screw was as effective as utilizing two screws to support the calcar. On the other hand, the variation in calcar screw configuration had minor influence on the fixation stability when all the proximal screws (A-D level) were filled. This study evaluates different screw configurations to further understand the influence of combined screw configurations and the individual screws on the fixation stability. Findings from this study may help decrease the risk for screw cut-out with proximal humerus varus collapse and the associated economic costs.
publisherThe American Society of Mechanical Engineers (ASME)
titleFinite Element-Predicted Effects of Screw Configuration in Proximal Humerus Fracture Fixation
typeJournal Paper
journal volume142
journal issue8
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.4045907
page81005
treeJournal of Biomechanical Engineering:;2020:;volume( 142 ):;issue: 008
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)