description abstractMost blood-contacting medical devices must be assessed for potential thrombogenicity prior to regulatory approval. A common assay for screening and qualifying devices involves monitoring the reduction of platelet and leukocyte (P&L) counts in whole blood exposed to the device. We have validated an improved method for assessing a device's effect on platelet activation and surface adhesion, offering significant improvement over the current ASTM F2888-13 method, which uses blood fully anticoagulated by acidified citrate (known to significantly inhibit platelet responsiveness). Our method uses minimal heparinization (final concentration 1 IU/mL) to optimize the response to commonly used control materials: latex, black rubber, and high-density polyethylene (HDPE). We also have shown the assay's capacity to appropriately assess a legally marketed comparator device (LMCD) with a documented clinical history. The test materials were prepared for incubation and allowed to remain in contact with the citrated or heparinized blood for ∼1 h at 37 °C. A complete blood count was performed prior to exposure, and at the end of the incubation period, reductions in P&L counts were recorded. Results from citrate-anticoagulated assay showed only a marginal response to the positive control, black rubber. Using heparinized blood, the assay generated a robust response to the positive controls, the “intermediate scoring” controls, and also assessed a legally marketed and approved device as clearly nonthrombogenic. This modification adds robustness and sensitivity to this quick and inexpensive thrombogenicity assay and should be incorporated into the next ASTM standards.
publisherThe American Society of Mechanical Engineers (ASME)
titleImproved Method for the ASTM Platelet and Leukocyte Assay: Use of Minimal Heparinization in a Screening Test for Hemocompatibility of Blood-Contacting Medical Devices
typeJournal Paper
journal volume13
journal issue1
journal titleJournal of Medical Devices
identifier doi10.1115/1.4041805
journal fristpage11004
journal lastpage011004-7
treeJournal of Medical Devices:;2019:;volume( 013 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record