Show simple item record

contributor authorValdovinos, John
contributor authorChris Bouwmeester, J.
contributor authorBonde, Pramod
date accessioned2019-02-28T11:04:59Z
date available2019-02-28T11:04:59Z
date copyright11/9/2017 12:00:00 AM
date issued2018
identifier issn1932-6181
identifier othermed_012_01_011001.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252485
description abstractRight ventricular (RV) dysfunction has limited the effectiveness of mechanical circulatory support (MCS) therapy in some heart failure (HF) patients. Intravascular pumps can provide adequate circulatory support without the need for extensive operations. The development of an intravascular right ventricular assist device (RVAD), called the cavo-arterial pump (CAP), is presented. Two prototypes of the CAP were developed to demonstrate the feasibility of providing adequate pulmonary support and to demonstrate the feasibility of using axial magnetic couplings for contactless torque transmission from the motor shaft to the pump impeller. The CAP utilizing a direct drive mechanism produced a maximum pressure of 100 mm Hg and a maximum flow of 2.25 L/min when operated at 24 kRPM. When a magnetic drive mechanism was used, the overall flowrate decreased due to a loss in torque transmission. The magnetic drive CAP was able to operate up to 18.5 kRPM and produce a maximum flowrate of 1.35 L/min and a maximum pressure difference of 40 mm Hg. These results demonstrate that the CAP produces sufficient output for partial circulatory support of the pulmonary circulation, and that axial magnetic couplings can help to eliminate the sealing system needed to isolate the miniature motor and bearings from blood contact.
publisherThe American Society of Mechanical Engineers (ASME)
titlePreliminary Design and Testing of a Cavo-Arterial Pump Utilizing Axial Magnetic Couplings
typeJournal Paper
journal volume12
journal issue1
journal titleJournal of Medical Devices
identifier doi10.1115/1.4038221
journal fristpage11001
journal lastpage011001-7
treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record