Show simple item record

contributor authorSoenksen, Luis R.
contributor authorMartínez-Corona, David A.
contributor authorIñiguez de Gante, Sofía
contributor authorPhabmixay, Pierre S.
contributor authorMarongiu Maggi, Mauricio J.
date accessioned2019-02-28T11:04:58Z
date available2019-02-28T11:04:58Z
date copyright1/30/2018 12:00:00 AM
date issued2018
identifier issn1932-6181
identifier othermed_012_01_011009.pdf
identifier urihttp://yetl.yabesh.ir/yetl1/handle/yetl/4252482
description abstractThe shelf life of point-of-care and rapid diagnostic tests (POC-RDTs) is commonly compromised by abrupt temperature changes during storage, transportation, and use. This situation is especially relevant in tropical regions and resource-constrained settings where cold chain may be unreliable. Here, we report the use of novel and low-cost passive thermal shield (TS) made from laminated phase change material (PCM) to reduce thermal overload in POC-RDTs. Validation of the proposed design was done through numerical simulation and testing of an octadecane shield prototype in contact with a lateral flow immunoassay. The use of our TS design provided 30–45 min delay in thermal equilibration under constant and oscillating heat load challenges resembling those of field use. The addition of a thin PCM protection layer to POC-RDTs can be a cost-effective, scalable, and reliable solution to provide additional thermal stability to these devices.
publisherThe American Society of Mechanical Engineers (ASME)
titleLow-Cost Thermal Shield for Rapid Diagnostic Tests Using Phase Change Materials
typeJournal Paper
journal volume12
journal issue1
journal titleJournal of Medical Devices
identifier doi10.1115/1.4038898
journal fristpage11009
journal lastpage011009-12
treeJournal of Medical Devices:;2018:;volume( 012 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record