| contributor author | Laan, Danuel | |
| contributor author | Diem Vu, Trang Ngoc | |
| contributor author | Hernandez, Matthew | |
| contributor author | Schiller, Henry | |
| date accessioned | 2019-02-28T11:04:50Z | |
| date available | 2019-02-28T11:04:50Z | |
| date copyright | 3/5/2018 12:00:00 AM | |
| date issued | 2018 | |
| identifier issn | 1932-6181 | |
| identifier other | med_012_02_025001.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl1/handle/yetl/4252458 | |
| description abstract | Tube thoracostomy (TT) insertion can serve as a life-saving adjunct for thoracic trauma. Unfortunately, suboptimal positioning using the open, standard of care technique is associated with complications resulting in impaired TT function. Using a porcine model, we aimed to determine whether a magnetic chest tube positioning system (MCTPS) could be utilized to direct the intrathoracic TT position. Using recently deceased cross-bred domestic swine, we performed TT using our MCTPS and the standard of care open technique. The operator held one magnet outside of the thorax. The second magnet was positioned at the distal aspect of the TT. The operator was tasked with positioning the TT to distinct premarked intra-thoracic locations under blinded conditions. The experiment was video-recorded through an open sternotomy incision. As a control, TT was inserted using the standard of care open technique. The utilization of MCTPS successfully directed TT from one premarked location to another in 4 of 5 attempts (80%). Conversely, the control TT without magnetic guidance failed to navigate the premarked intra-thoracic locations with 0 of 5 attempts successful (p = 0.05). Positional flaws after TT placement are common. We demonstrate the feasibility of the MCTPS as an alternative to traditional hand-guided technique under simulated TT insertion conditions. The MCTPS is possibly superior to the current standard of care technique of TT. Additional studies are needed to develop this emerging technology in humans. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Magnetic Chest Tube Positioning System | |
| type | Journal Paper | |
| journal volume | 12 | |
| journal issue | 2 | |
| journal title | Journal of Medical Devices | |
| identifier doi | 10.1115/1.4039208 | |
| journal fristpage | 25001 | |
| journal lastpage | 025001-3 | |
| tree | Journal of Medical Devices:;2018:;volume( 012 ):;issue: 002 | |
| contenttype | Fulltext | |