Show simple item record

contributor authorGrove, Kent
contributor authorDeline, Steve M.
contributor authorSchatz, Tim F.
contributor authorHoward, Sarah E.
contributor authorPorter, Deanna
contributor authorSmith, Mark E.
date accessioned2017-11-25T07:18:31Z
date available2017-11-25T07:18:31Z
date copyright2017/3/5
date issued2017
identifier issn1932-6181
identifier othermed_011_02_021008.pdf
identifier urihttp://138.201.223.254:8080/yetl1/handle/yetl/4235215
description abstractISO 10993-4 in vivo thrombogenicity testing is frequently performed for regulatory approval of many blood-contacting medical devices and is often a key part of submission packages. Given the current state of in vivo thrombogenicity assays, a more robust and reproducible assay design, including in vitro models, is needed. This study describes an in vitro assay that integrates freshly harvested ovine blood containing minimal heparin in a closed pumped loop. To confirm the reproducibility of this assay, control materials were identified that elicited either a positive or a negative thrombogenic response. These controls demonstrated reproducibility in the resulting thrombogenicity scores with median scores of 5 and 0 for the positive and negative controls, respectively, which also demonstrated a significant difference (p < 0.0001). For a direct comparison of the in vitro blood loop assay to the traditional in vivo nonanticoagulated venous implant (NAVI) assay, seven sheep were used as blood donors for the loop and then as subjects for an NAVI assay. In each assay—loop or NAVI—three study articles were used: the positive and negative controls and a marketed, approved catheter. The resulting thrombogenicity scores were similar when comparing the loop to the NAVI results. For each study article, the median thrombogenicity scores were the same in these two different assays, being 0, 1, and 5 for the negative control, the marketed catheter, and the positive control, respectively. These data suggest that the in vitro assay performs similarly to the in vivo NAVI assay. This in vitro blood loop method has the potential to predict a materials' in vivo thrombogenicity, can substantially de-risk the materials or coating selection process, and may eventually be able to replace the in vivo models currently in use.
publisherThe American Society of Mechanical Engineers (ASME)
titleThrombogenicity Testing of Medical Devices in a Minimally Heparinized Ovine Blood Loop
typeJournal Paper
journal volume11
journal issue2
journal titleJournal of Medical Devices
identifier doi10.1115/1.4035724
journal fristpage21008
journal lastpage021008-8
treeJournal of Medical Devices:;2017:;volume( 011 ):;issue: 002
contenttypeFulltext