Show simple item record

contributor authorWishart, Edward R.
date accessioned2017-06-09T17:39:42Z
date available2017-06-09T17:39:42Z
date copyright1978/10/01
date issued1978
identifier issn0021-8952
identifier otherams-9570.pdf
identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4233073
description abstractBy optimization of the signal-to-noise ratio of a negative-feedback charge-sensitive electrometer, a noise level of about 0.7 fC peak to peak has been obtained. Output signal pulse width above the 1% amplitude level is 50 ms. Five parallel final amplifiers produce ±10 V outputs for input charges from ±1 fC to ±10 pC.
publisherAmerican Meteorological Society
titleAn Ultra-Sensitive Instrument for Measuring Electric Charges on Ice Particles
typeJournal Paper
journal volume17
journal issue10
journal titleJournal of Applied Meteorology
identifier doi10.1175/1520-0450(1978)017<1510:AUSIFM>2.0.CO;2
journal fristpage1510
journal lastpage1513
treeJournal of Applied Meteorology:;1978:;volume( 017 ):;issue: 010
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)