Show simple item record

contributor authorBasher, Reid E.
contributor authorMatthews, W. Andrew
date accessioned2017-06-09T17:39:07Z
date available2017-06-09T17:39:07Z
date copyright1977/08/01
date issued1977
identifier issn0021-8952
identifier otherams-9306.pdf
identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4232780
description abstractAn analysis of the use of ultraviolet narrow-band interference filters for total ozone determination is given with reference to the New Zealand filter spectrophotometer under the headings of filter monochromaticity, temperature dependence, orientation dependence, aging, and specification tolerances and non-uniformity. Quantitative details of each problem are given, together with the means used to overcome them in the New Zealand instrument. The tuning of the instrument's filter center wavelengths to a common set of values by tilting the filters is also described, along with a simple calibration method used to adjust and set these center wavelengths.
publisherAmerican Meteorological Society
titleProblems in the Use of Interference Filters for Spectrophotometric Determination of Total Ozone
typeJournal Paper
journal volume16
journal issue8
journal titleJournal of Applied Meteorology
identifier doi10.1175/1520-0450(1977)016<0795:PITUOI>2.0.CO;2
journal fristpage795
journal lastpage802
treeJournal of Applied Meteorology:;1977:;volume( 016 ):;issue: 008
contenttypeFulltext