Show simple item record

contributor authorHovenac, Edward A.
contributor authorHirleman, E. Dan
date accessioned2017-06-09T16:13:36Z
date available2017-06-09T16:13:36Z
date copyright1991/02/01
date issued1991
identifier issn0739-0572
identifier otherams-637.pdf
identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4204733
description abstractCalibration devices for the Forward Scattering Spectrometer Probe (FSSP) and the Optical Array Probe (OAP) were developed. The device used with the FSSP is a rotating pinhole calibrator. It utilizes light diffracted by a pinhole of a diameter to simulate scattered light from a water droplet. This device can be used to calibrate the FSSP, measure the FSSP's optical collection angles and for instrument alignment and troubleshooting. The device used with the OAP is a rotating reticle calibrator. Chrome disks of a known diameter on the reticle are used for calibration of the OAP and for determining the OAP's response to our-of-focus particles in the probe volume.
publisherAmerican Meteorological Society
titleUse of Rotating Pinholes and Reticles for Calibration of Cloud Droplet Instrumentation
typeJournal Paper
journal volume8
journal issue1
journal titleJournal of Atmospheric and Oceanic Technology
identifier doi10.1175/1520-0426(1991)008<0166:UORPAR>2.0.CO;2
journal fristpage166
journal lastpage171
treeJournal of Atmospheric and Oceanic Technology:;1991:;volume( 008 ):;issue: 001