Show simple item record

contributor authorPade, Gerald
contributor authorCovert, David
contributor authorLarson, Timothy
date accessioned2017-06-09T14:54:59Z
date available2017-06-09T14:54:59Z
date copyright1987/06/01
date issued1987
identifier issn0739-0572
identifier otherams-296.pdf
identifier urihttp://onlinelibrary.yabesh.ir/handle/yetl/4166844
description abstractAn inertial impactor is described that is designed to actively collect fog droplets from a flowing airstream. This collector samples 60 liters of air per second with a theoretical d50 currently set at 8 ?m. Collector performance was assessed by generating a known number of 22 and 44 ?m water droplets tagged with 1 ppm sodium fluorescein and injecting these droplets into a larger flow of polydisperse, untagged droplets immediately upstream of the collector. These large, chemically tagged droplets were found to be collected with 90% efficiency, in good agreement with theoretical estimates of collector performance. Between 3 and 8% water loss via evaporation was observed for relative humidities between 93% and 88%, respectively.
publisherAmerican Meteorological Society
titleDesign and Calibration of a High Volume Fogwater Collector
typeJournal Paper
journal volume4
journal issue2
journal titleJournal of Atmospheric and Oceanic Technology
identifier doi10.1175/1520-0426(1987)004<0340:DACOAH>2.0.CO;2
journal fristpage340
journal lastpage344
treeJournal of Atmospheric and Oceanic Technology:;1987:;volume( 004 ):;issue: 002