Show simple item record

contributor authorL. T. Akeley
date accessioned2017-05-09T01:30:10Z
date available2017-05-09T01:30:10Z
date copyrightDecember, 1972
date issued1972
identifier issn0098-2202
identifier otherJFEGA4-27401#893_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/161534
description abstractThe author reviews methods of calibrating low-differential (25–300 in. H2 O) D.P. transmitters at high (up to 3000 psig) static pressure. Various test methods are described along with discussion of their merits. Included are use of one or more deadweight testers (gas or liquid) and use of a high-pressure mercury manometer as a standard. Five D.P. transmitters of different construction are tested and their static span-shift performance reported.
publisherThe American Society of Mechanical Engineers (ASME)
titleDifferential Pressure Transmitters, Span-Shift Errors With Static Pressure
typeJournal Paper
journal volume94
journal issue4
journal titleJournal of Fluids Engineering
identifier doi10.1115/1.3425583
journal fristpage893
journal lastpage896
identifier eissn1528-901X
keywordsPressure
keywordsPressure sensors
keywordsErrors
keywordsHigh pressure (Physics)
keywordsManometers AND Construction
treeJournal of Fluids Engineering:;1972:;volume( 094 ):;issue: 004