description abstractThe paper describes measurement and calculation of 2D distribution of the vibration signal originated by the gradient coil system of the magnetic resonance imaging (MRI) equipment. The measurement experiments were performed at the bottom plastic holder of the scanning area of the openair weak magnetic field MRI device. Selection of a usable type of a vibration sensor for measurement in a magnetic field with low B0 up to 0.2 T is also discussed. Realized calibration of the chosen sensor (sensitivity and frequency response) together with determination of the propagation time delay between the excitation impulses and the subsequently generated vibration signal is mentioned, too. The pickedup vibration signal exhibits harmonic character so it is suitable to describe it by determined spectral properties and features. Obtained statistical results of spectral analysis will be used to improve image sharpening and reduction of the motion effect in the MR pictures of thin layer samples and phantoms scanned by this MRI system.
publisherThe American Society of Mechanical Engineers (ASME)
titleMapping and Spectral Analysis of Acoustic Vibration in the Scanning Area of the Weak Field Magnetic Resonance Imager
typeJournal Paper
journal volume136
journal issue5
journal titleJournal of Vibration and Acoustics
identifier doi10.1115/1.4027791
journal fristpage51005
journal lastpage51005
identifier eissn1528-8927
treeJournal of Vibration and Acoustics:;2014:;volume( 136 ):;issue: 005
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record