Show simple item record

contributor authorStacy L. Figueredo
contributor authorAlexander H. Slocum
date accessioned2017-05-09T00:40:01Z
date available2017-05-09T00:40:01Z
date copyrightJune, 2010
date issued2010
identifier issn1932-6181
identifier otherJMDOA4-28010#027515_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/144421
description abstractIn order to test small scale medical devices, it is often necessary to prototype them on a larger scale such that proof-of-concept tests can be made more affordably and design details can be tested more easily. This article discusses a method of scaling large needle prototypes for testing in a gelatin phantom such that puncture forces match those expected for the actual-size needle when puncturing tissue. Using Hertz contact force equations to account for the differences in prototype materials and size, as well as for the tissue phantom properties, 10× scale prototype needles were inserted into a gelatin phantom and puncture forces were compared with those of a real scale prototype in bovine liver tissue. Results showed that for a 19 gauge (1.06 mm) stainless steel needle tip, where a maximum doctor-applied load of 5 N was desired to pierce liver tissue, loads of 0.44 N using Hertz point contact, and 0.31 N using Hertz line contact methods were predicted to puncture liver tissue, and an average load of 0.31 N was observed in force-displacement measurements. With a 10× scale stereolithographed needle prototype, Hertz point contact predicted a load of 0.31 N to puncture a gelatin phantom, Hertz line contact theory predicted 0.37 N, and an average load of 0.73 N observed in force displacement measurements. Similar contact mechanics based scaling methods might be applied to cutting devices.
publisherThe American Society of Mechanical Engineers (ASME)
titleHertz Contact Based Scaling of Puncture Forces for Large Scale Needle Prototypes Using a Gelatinous Tissue Phantom
typeJournal Paper
journal volume4
journal issue2
journal titleJournal of Medical Devices
identifier doi10.1115/1.3442791
journal fristpage27515
identifier eissn1932-619X
treeJournal of Medical Devices:;2010:;volume( 004 ):;issue: 002
contenttypeFulltext