Show simple item record

contributor authorAmy B. Zavatsky
contributor authorAndrew J. Price
contributor authorPaul T. Oppold
date accessioned2017-05-09T00:12:19Z
date available2017-05-09T00:12:19Z
date copyrightJune, 2004
date issued2004
identifier issn0148-0731
identifier otherJBENDY-26369#351_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/129614
description abstractThis study involved the development and testing of a system for the simultaneous in vitro measurement of tibiofemoral kinematics and patellofemoral kinematics and forces. Knee motion was tracked using a Vicon 370, and patellofemoral force was measured using a six degree-of-freedom transducer based on the design of Singerman et al. Using this system, twelve knee specimens were tested in supine leg extension under a simulated quadriceps force. The comprehensive set of results corresponds well to the individual results of previous studies. The measurement system will be of value in assessing the effects of total knee arthroplasty on patellar function.
publisherThe American Society of Mechanical Engineers (ASME)
titleSimultaneous In Vitro Measurement of Patellofemoral Kinematics and Forces
typeJournal Paper
journal volume126
journal issue3
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.1762896
journal fristpage351
journal lastpage356
identifier eissn1528-8951
keywordsKinematics
keywordsForce
keywordsTransducers
keywordsKnee AND Motion
treeJournal of Biomechanical Engineering:;2004:;volume( 126 ):;issue: 003