Show simple item record

contributor authorB. Gasser
contributor authorB. Boman
contributor authorD. Wyder
contributor authorE. Schneider
date accessioned2017-05-08T23:32:07Z
date available2017-05-08T23:32:07Z
date copyrightFebruary, 1990
date issued1990
identifier issn0148-0731
identifier otherJBENDY-25855#15_1.pdf
identifier urihttp://yetl.yabesh.ir/yetl/handle/yetl/106603
description abstractIlizarov proposes the use of a special circular fixation device for treatment of bone defects and nonunions or for limb lengthening. Supposition was that the success of this method has to do with the specific mechanical behavior of the device. This behavior is a result of the configuration of the fixation elements. Therefore, a mechanical study of the sensitivity of the circular compression and distraction device (CDD) to configuration parameters was performed. The CDD was found to exhibit a nonlinear stiffness behavior, in particular under axial load. This may be favorable for the induction and tolerance of bone formation. Among the different parameters tested the ring radius was the most important with respect to stiffness. In general, the stiffness of the CDD allows adjustment during postoperative management. In magnitude, it is equal or lower when compared to other fixator types.
publisherThe American Society of Mechanical Engineers (ASME)
titleStiffness Characteristics of the Circular Ilizarov Device as Opposed to Conventional External Fixators
typeJournal Paper
journal volume112
journal issue1
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.2891120
journal fristpage15
journal lastpage21
identifier eissn1528-8951
keywordsStiffness
keywordsBone
keywordsMechanical behavior
keywordsCompression
keywordsElectromagnetic induction
keywordsProduct quality AND Stress
treeJournal of Biomechanical Engineering:;1990:;volume( 112 ):;issue: 001
contenttypeFulltext


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record