| contributor author | B. Gasser | |
| contributor author | B. Boman | |
| contributor author | D. Wyder | |
| contributor author | E. Schneider | |
| date accessioned | 2017-05-08T23:32:07Z | |
| date available | 2017-05-08T23:32:07Z | |
| date copyright | February, 1990 | |
| date issued | 1990 | |
| identifier issn | 0148-0731 | |
| identifier other | JBENDY-25855#15_1.pdf | |
| identifier uri | http://yetl.yabesh.ir/yetl/handle/yetl/106603 | |
| description abstract | Ilizarov proposes the use of a special circular fixation device for treatment of bone defects and nonunions or for limb lengthening. Supposition was that the success of this method has to do with the specific mechanical behavior of the device. This behavior is a result of the configuration of the fixation elements. Therefore, a mechanical study of the sensitivity of the circular compression and distraction device (CDD) to configuration parameters was performed. The CDD was found to exhibit a nonlinear stiffness behavior, in particular under axial load. This may be favorable for the induction and tolerance of bone formation. Among the different parameters tested the ring radius was the most important with respect to stiffness. In general, the stiffness of the CDD allows adjustment during postoperative management. In magnitude, it is equal or lower when compared to other fixator types. | |
| publisher | The American Society of Mechanical Engineers (ASME) | |
| title | Stiffness Characteristics of the Circular Ilizarov Device as Opposed to Conventional External Fixators | |
| type | Journal Paper | |
| journal volume | 112 | |
| journal issue | 1 | |
| journal title | Journal of Biomechanical Engineering | |
| identifier doi | 10.1115/1.2891120 | |
| journal fristpage | 15 | |
| journal lastpage | 21 | |
| identifier eissn | 1528-8951 | |
| keywords | Stiffness | |
| keywords | Bone | |
| keywords | Mechanical behavior | |
| keywords | Compression | |
| keywords | Electromagnetic induction | |
| keywords | Product quality AND Stress | |
| tree | Journal of Biomechanical Engineering:;1990:;volume( 112 ):;issue: 001 | |
| contenttype | Fulltext | |