description abstractCompliant vascular grafts were modeled by finite elasticity theory. A linear, biaxial model satisfactorily described the stress-strain behavior in inflation tests, where the sample length was fixed longitudinally and inflated. The model was then used to predict the behavior in a longitudinal test (where the longitudinal stretch was varied while keeping the pressure zero), and in a uniaxial test of a circumferential strip. The model satisfactorily predicted the longitudinal Young’s modulus measured in the longitudinal test. The model was less successful in predicting the circumferential Young’s modulus measured in the uniaxial test, possibly because the state of stress was not purely uniaxial.
publisherThe American Society of Mechanical Engineers (ASME)
titleFinite Elasticity Modeling of the Biaxial and Uniaxial Properties of Compliant Vascular Grafts
typeJournal Paper
journal volume110
journal issue4
journal titleJournal of Biomechanical Engineering
identifier doi10.1115/1.3108451
journal fristpage344
journal lastpage348
identifier eissn1528-8951
keywordsElasticity
keywordsModeling
keywordsStress
keywordsPressure
keywordsInflationary universe AND Strips
treeJournal of Biomechanical Engineering:;1988:;volume( 110 ):;issue: 004